ARJOHUNTLEIGH POLSKA Sp. z.o.o. recalls Arjo Tenor mobile passive patient lift
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- An internal component of the actuators in a certain production batch may not meet the required mechanical strength. This could result, in the worst case scenario, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm.
- Products
-
- Arjo Tenor mobile passive patient lift
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
05055982787758- Units affected
- 119 units
- About this recall
- Arjo Tenor mobile passive patient lift, Model Numbers KHA1000 US and KHA1010 US