Philips Respironics, Inc. recalls BiPAP V30 Auto Ventilator
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential for devices to experience interruption/loss of therapy during a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.
- Products
-
- BiPAP V30 Auto Ventilator
- Company
- Philips Respironics, Inc.
- Where it was sold
-
Worldwide distribution. US Nationwide including PR and GU; Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hungary, Italy, Netherlands, Poland, Portugal, Romania, Serbia, Slovakia, Spain, and Turkey.
Nationwide
- Barcodes
00606959049635- Units affected
- 5,958 US; None OUS
- About this recall
- BiPAP V30 Auto Ventilator, Part Number 1111178