Onkos Surgical, Inc. recalls ELEOS COLLAR STEM
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.
- Products
-
- ELEOS COLLAR STEM
- Company
- Onkos Surgical, Inc.
- Where it was sold
-
US Nationwide distribution in the states of California, Kentucky, and Wisconsin.
States: California, Kentucky, Wisconsin
- Units affected
- 23 units
- About this recall
- ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.