Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America Llc recalls Achieva 1.5T Initial system Model Number (REF): 781178

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Products
  • Achieva 1.5T Initial system Model Number (REF): 781178
Where it was sold

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Nationwide

Units affected
2 units
About this recall
Achieva 1.5T Initial system Model Number (REF): 781178