Philips North America Llc recalls Ingenia 1.5T S Model Number (REF): 781347
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
- Products
-
- Ingenia 1.5T S Model Number (REF): 781347
- Company
- Philips North America Llc
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Nationwide
- Barcodes
00884838068421- Units affected
- 139 units
- About this recall
- Ingenia 1.5T S Model Number (REF): 781347