Philips North America Llc recalls MR RT Upgrades- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnost
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
- Products
-
- MR RT Upgrades- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnost
- Company
- Philips North America Llc
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Nationwide
- Units affected
- 2 units
- About this recall
- MR RT Upgrades- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781440