Silex Medical, LLC recalls Signature Laparoscopic Instruments
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Products
-
- Signature Laparoscopic Instruments
- Company
- Silex Medical, LLC
- Where it was sold
-
US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.
States: California, Connecticut, Florida, Hawaii, Iowa, Idaho, Illinois, Michigan, Mississippi, North Dakota, New Jersey, Oregon, Texas, Virginia, West Virginia
- Units affected
- 105 units
- About this recall
- Signature Laparoscopic Instruments, Atraumatic Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5864000.