Lucid Diagnostics, Inc. recalls Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.
- Products
-
- Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter
- Company
- Lucid Diagnostics, Inc.
- Where it was sold
-
US Nationwide distribution in the states of MN, MD.
States: Maryland, Minnesota
- Barcodes
00860007510977- Units affected
- 220 units (US)
- About this recall
- Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter, 10 pores/inch) Model/Catalog Number: EC25T10VI Software Version: N/A Product Description: EsophaCap is pouched as individual units. 10 units go into a shelf-box. Component: N/A