Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Lucid Diagnostics, Inc. recalls Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper esophageal sphincter between the pharynx and the esophagus, which is the narrowest part of the alimentary canal.
Products
  • Brand Name: EsophaCap Product Name: EsophaCap (25mm diameter
Where it was sold

US Nationwide distribution in the states of MN, MD.

States: Maryland, Minnesota

Barcodes
00860007510977
Units affected
220 units (US)
About this recall