Beckman Coulter, Inc. recalls Access Testosterone assay

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
Products
  • Access Testosterone assay
Where it was sold

US State: FL. UAE, Vietnam

States: Florida

Barcodes
15099590231347
Units affected
70 units
About this recall
Access Testosterone assay, Catalog Number 33560