Beckman Coulter, Inc. recalls Access Testosterone assay
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.
- Products
-
- Access Testosterone assay
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
US State: FL. UAE, Vietnam
States: Florida
- Barcodes
15099590231347- Units affected
- 70 units
- About this recall
- Access Testosterone assay, Catalog Number 33560