Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Dexcom, Inc. recalls Dexcom G7 iOS Application

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.
Products
  • Dexcom G7 iOS Application
Where it was sold

Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: United Arab Emirates (AE), Austria (AT), Australia (AU), Belgium (BE), Bahrain (BH), Canada (CA), Switzerland (CH), Czech Republic (CZ), Germany (DE), Denmark (DK), Spain (ES), Finland (FI), France (FR), United Kingdom (GB), Hong Kong (HK), Ireland (IE), Israel (IL), Iceland (IS), Italy (IT), Jordan (JO), Korea, Repulic (KR), Kuwait (KW), Lebanon (LB), Lithuania (LT), Luxemburg (LU), Malta (MT), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Poland (PL), Portugal (PT), Qatar (QA), Saudi Arabia (SA), Sweden (SE), Slovenia (SI), Slovakia (SK), Turkey (TR) and South Africa (ZA).

Nationwide

Barcodes
00386270001863
Units affected
61,355 users
About this recall
Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system