American Contract Systems, Inc. recalls LOWER EXTREMITY
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
- Products
-
- LOWER EXTREMITY
- Where it was sold
-
US Nationwide distribution in the state of CO.
States: Colorado
- Units affected
- 850 Kits
- About this recall