Remel Europe Ltd. recalls Oxoid Agglutinating Sera
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
- Products
-
- Oxoid Agglutinating Sera
- Company
- Remel Europe Ltd.
- Where it was sold
-
Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
Nationwide
- Barcodes
05056080500270- Units affected
- 117 US; 950 OUS
- About this recall
- Oxoid Agglutinating Sera, Salmonella 9-0 R30957301