Medtronic, Inc. recalls Azure S DR MRI SureScan
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Identified devices may experience sudden battery power failure without alarm, power supply power-on-resets, and interruption of therapy. Potential patient harms include insufficient pacing support leading to cardiac arrest, syncope, or hospitalization; or system revision (explant/replacement) leading to infection, hematoma, or temporary impairment.
- Products
-
- Azure S DR MRI SureScan
- Company
- Medtronic, Inc.
- Where it was sold
-
International distribution of the country of UK. No US distribution.
- Barcodes
00763000956004- Units affected
- 6
- About this recall
- Azure S DR MRI SureScan, Product number W3DR01