Philips North America recalls Intera 1.5T Achieva Nova Product Number: 781172;
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
- Products
-
- Intera 1.5T Achieva Nova Product Number: 781172;
- Company
- Philips North America
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
States: Arizona, California, Delaware, Massachusetts, Michigan, Nebraska, New York, South Dakota
- Units affected
- 16 units (1 US, 15 OUS)
- About this recall
- Intera 1.5T Achieva Nova Product Number: 781172;