Baxter Healthcare Corporation recalls Baxter Novum IQ Syringe Pump
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.
- Products
-
- Baxter Novum IQ Syringe Pump
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
05413765852428- Units affected
- 8027 units
- About this recall
- Baxter Novum IQ Syringe Pump, product code 40800BAXUS,