Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer C11137
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.
- Products
-
- DxI 9000 Access Immunoassay Analyzer C11137
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI, ID, IL, IN, KY, LA, ME, MI, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, WA, WI and the countries of Algeria, Australia, Austria, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, France, Germany, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Korea, Republic of, Libya, New Zealand, Oman, Poland, Portugal, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.
States: Arkansas, California, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Kentucky, Louisiana, Maine, Michigan, New Hampshire, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Washington, Wisconsin
- Barcodes
15099590732103- Units affected
- 67 US; 178 OUS
- About this recall
- DxI 9000 Access Immunoassay Analyzer C11137