Maquet Critical Care AB recalls Servo-u Ventilator System. Model Number: 6688600.
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
- Products
-
- Servo-u Ventilator System. Model Number: 6688600.
- Company
- Maquet Critical Care AB
- Where it was sold
-
US Nationwide distribution and Puerto Rico. No international distribution.
Nationwide
- Barcodes
07325710001042- Units affected
- 874 units
- About this recall
- Servo-u Ventilator System. Model Number: 6688600.