Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ge Medical Systems Israel, Functional Imaging recalls MyoSPECT System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.
Products
  • MyoSPECT System
Where it was sold

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, NC, ND, NV, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Belgium, Brazil, Canada, Czech Republic, Finland, France, India, Israel, Italy, Japan, Malaysia, Malta, Poland, Singapore, Spain, Taiwan, Turkey, and United Kingdom.

States: Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Iowa, Illinois, Indiana, Louisiana, Massachusetts, Maryland, Michigan, Missouri, North Carolina, North Dakota, Nevada, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Virginia, Washington

Barcodes
00195278421586001952784886190019527842157900195278488626
Units affected
133 devices
About this recall
MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).