Ge Medical Systems Israel, Functional Imaging recalls Varicam
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
- Products
-
- Varicam
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 21
- About this recall
- Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System