Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Products
  • Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
10195327331795101953273189251019532752481410198459085925
Units affected
892 units
About this recall
Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY, Model Number: DYNDA2941A; 2) CHEST/BREAST-LF, Model Number: DYNJ83950; 3) HERNIA REPAIR TOTE, Model Number: DYNJ908337A; 4) HERNIA REPAIR TOTE, Model Number: DYNJ908337C