Medline Industries, LP recalls Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Products
-
- Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
10195327331795101953273189251019532752481410198459085925- Units affected
- 892 units
- About this recall
- Medline Convenience Kits: 1) CHEST TUBE INSERTION TRAY, Model Number: DYNDA2941A; 2) CHEST/BREAST-LF, Model Number: DYNJ83950; 3) HERNIA REPAIR TOTE, Model Number: DYNJ908337A; 4) HERNIA REPAIR TOTE, Model Number: DYNJ908337C