Medline Industries, LP recalls Medline Convenience Kits: 1) NEURO SHUNT- CODMAN
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Products
-
- Medline Convenience Kits: 1) NEURO SHUNT- CODMAN
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
10889942459643101953273817761019348947815010193489478198101953272407071019348947836510195327240776101934894776341019532724110010195327110932101953275729141019532711585210193489301342101934899191961019532763873310193489712100101934893352481088994244455710193489502053101934898854221019348922553210195327126964101953274702341019532760591910195327690984101953276060771019348942271910193489615012101953272428001019348944481010195327516338101953274436891019532751856110195327579289101984590321651019845916612910198459233159101984595612451019532769741910193489906615101953272902691019845927255410198459408519108899421175291019348998541210195327067595101953274904161019845911892010195327404741101953276615951019845909639610193489333824- Units affected
- 10184 units
- About this recall
- Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) NEURO/LAMINECTOMY BASIN PK-LF, Model Number: DYNJ0781823K; 15) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 16) NEURO PACK, Model Number: DYNJ49098I; 17) NEURO PACK, Model Number: DYNJ53089B; 18) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 19) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 20) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 21) NEURO PACK, Model Number: DYNJ60792B; 22) NEURO PACK, Model Number: DYNJ60792C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 24) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 25) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157F; 26) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 27) NEURO PACK, Model Number: DYNJ61740D; 28) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 29) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 30) DISCECTOMY PACK, Model Number: DYNJ67835; 31) FSC NEURO PACK, Model Number: DYNJ80442B; 32) WWD NEURO PACK, Model Number: DYNJ80849B; 33) WWD NEURO PACK, Model Number: DYNJ80849C; 34) WWD NEURO PACK, Model Number: DYNJ80849D; 35) WWD NEURO PACK, Model Number: DYNJ80849F; 36) WWD NEURO PACK, Model Number: DYNJ80849G; 37) WWD NEURO PACK, Model Number: DYNJ80849I; 38) WWD NEURO PACK, Model Number: DYNJ80849J; 39) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 40) NEURO, Model Number: DYNJ900904J; 41) NEURO, Model Number: DYNJ900904K; 42) NEURO, Model Number: DYNJ900904L; 43) NEURO, Model Number: DYNJ900904M; 44) NEURO, Model Number: DYNJ902585; 45) NEURO, Model Number: DYNJ902585A; 46) NEURO - IFCAP 230316, Model Number: DYNJ908393; 47) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 48) NEURO - IFCAP 230316, Model Number: DYNJ908393B; 49) NEURO KIT, Model Number: DYNJ909573; 50) NEURO, Model Number: DYNJ910384; 51) NEURO, Model Number: DYNJ910384A; 52) NEURO PACK-LF, Model Number: PHS390015M