Alcon Research LLC recalls Brand Name: Alcon Laboratories
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.
- Products
-
- Brand Name: Alcon Laboratories
- Company
- Alcon Research LLC
- Where it was sold
-
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Albania, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Ecuador, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Luxembourg, Macao, Malaysia, Malta, Mexico, Moldova, Mongolia, Montenegro, Morocco, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Qatar, Republic of North Macedonia, R¿union, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.
Nationwide
- Barcodes
380650000934380650000958380650000965380657524358380657524365380657524372380657524389380657524396380657524495380657524501380657524518380657531066380657531097380658300777- Units affected
- 198205
- About this recall
- Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Vitrectomy Procedure Pak Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' Software Version: N/A Product Description: Surgical procedure pack to interface with Alcon Constellation Surgical console. Component: N/A