Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Products
  • Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
101934894788391019348989789010195327534141
Units affected
753 units
About this recall
Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55A; 3) TRACH TOTE, Model Number:DYNJ85691