Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).
Products
  • VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides)
Where it was sold

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Nationwide

Units affected
414,000 (OUS)
About this recall
VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);