FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline Convenience Kits: 1) LVAD DRIVELINE TRAY

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Products
  • Medline Convenience Kits: 1) LVAD DRIVELINE TRAY
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
1065316032775210653160345251653160177473
Units affected
9720 units
About this recall
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A