Aesculap Inc recalls Aeos Robotic Digital Microscope
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.
- Products
-
- Aeos Robotic Digital Microscope
- Company
- Aesculap Inc
- Where it was sold
-
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
States: Arkansas, Florida, Illinois, Indiana, Kentucky, Michigan, North Carolina, Nebraska, Oklahoma, South Dakota, Texas
- Barcodes
04046955206742- Units affected
- 19 units
- About this recall
- Aeos Robotic Digital Microscope, Product Code: PV010