Beckman Coulter Inc. recalls Estrone RIA
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
- Products
-
- Estrone RIA
- Company
- Beckman Coulter Inc.
- Where it was sold
-
US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
States: North Carolina
- Barcodes
15099590211615- Units affected
- 496 units
- About this recall
- Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.