Medtronic, Inc. recalls Medtronic CareLink SmartSync Device Manager
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.
- Products
-
- Medtronic CareLink SmartSync Device Manager
- Company
- Medtronic, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
0076300000203900763000002046007630000020530076300039785200763000397869007630003978760076300039788300763000397890007630003979060076300039791300763000544300- Units affected
- 22091 total devices
- About this recall
- Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;