Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Dental EZ Group Star Dental Division recalls Titan Blis-Sonic Scaler packaged with the Titan Scaler Tip - Perio (part number 261669) (1) P/N 264563 (2) P/N

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Tip holder portion of the scaler tip assembly may have an oversized condition that will not allow the scaler tip to stay seated in the holder may cause injury
Products
  • Titan Blis-Sonic Scaler packaged with the Titan Scaler Tip - Perio (part number 261669) (1) P/N 264563 (2) P/N
Where it was sold

US Nationwide distribution.

Nationwide

Units affected
192 units
About this recall
Titan Blis-Sonic Scaler packaged with the Titan Scaler Tip - Perio (part number 261669) (1) P/N 264563 (2) P/N 264565 The scaler tip is used with an air drive scaler for subgingival scaling and fine access. The Perio tip is used for lingual supragingival scaling of mandibular incisors and deep local pockets.