Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medtronic Neuromodulation recalls Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
Products
  • Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
0064316989092300643169890947006431698909610076300007773000763000094027007630004210070076300044709000763000729141007630007372900076300092614400763000984649
Units affected
10165 units
About this recall
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump