Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
- Products
-
- LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
- Company
- Fresenius Kabi USA, LLC
- Where it was sold
-
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
States: California, Colorado, Florida, Georgia, Idaho, Illinois, Maryland, Michigan, Minnesota, Missouri, Mississippi, Nebraska, New Jersey, Nevada, Oklahoma, Oregon, South Carolina, Texas, Virginia, Washington, Wisconsin
- Barcodes
00811505030122- Units affected
- 32 systems
- About this recall
- LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.