Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS)

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Products
  • Ivenix Infusion System (IIS)
Where it was sold

Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;

States: California, Colorado, Florida, Georgia, Idaho, Illinois, Maryland, Michigan, Minnesota, Missouri, Mississippi, Nebraska, New Jersey, Nevada, Oklahoma, Oregon, South Carolina, Texas, Virginia, Washington, Wisconsin

Barcodes
00811505030320
Units affected
18,444
About this recall
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004