Philips Medical Systems (Cleveland) Inc recalls Pinnacle 3 with TumorLOC
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
- Products
-
- Pinnacle 3 with TumorLOC
- Where it was sold
-
Worldwide - US Nationwide distribution in the state of WV and the countries of Macao & United Kingdom.
States: West Virginia
- Barcodes
00884838103566- Units affected
- 149 Serial Numbers
- About this recall
- Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software