Beckman Coulter, Inc. recalls DxI 800 Access Immunoassay Analyzer W/Spot B
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
- Products
-
- DxI 800 Access Immunoassay Analyzer W/Spot B
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide.
Nationwide
- Barcodes
15099590369194- Units affected
- 451 systems
- About this recall
- DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456