Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Beckman Coulter, Inc. recalls DxI 600 Access Immunoassay Analyzer W/Dual Gantry

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Products
  • DxI 600 Access Immunoassay Analyzer W/Dual Gantry
Where it was sold

Worldwide distribution - US Nationwide.

Nationwide

Barcodes
15099590369248
Units affected
1 system
About this recall
DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461