GE Vingmed Ultrasound As recalls GE Vivid Ultrasound products
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
- Products
-
- GE Vivid Ultrasound products
- Company
- GE Vingmed Ultrasound As
- Where it was sold
-
US: IL, MA, NM, NY, OK,
States: Illinois, Massachusetts, New Mexico, New York, Oklahoma
- Units affected
- 7 units
- About this recall
- GE Vivid Ultrasound products, Model Numbers: a) Vivid E90, H45581LB; b) Vivid E95, H45581DA