Baxter Healthcare Corporation recalls Welch Allyn CP150 Electrocardiograph with Spirometry Option
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
- Products
-
- Welch Allyn CP150 Electrocardiograph with Spirometry Option
- Where it was sold
-
Distribution to Illinois.
States: Illinois
- Barcodes
00732094177640- Units affected
- 1 unit
- About this recall
- Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.