Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Pentax of America Inc recalls Pentax Medical Video Processor; Model Number: EPK-i8020c;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
Products
  • Pentax Medical Video Processor; Model Number: EPK-i8020c;
Where it was sold

US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.

States: Arizona, California, District of Columbia, Florida, Iowa, Idaho, Illinois, Kansas, Massachusetts, Maryland, Michigan, Missouri, North Carolina, New Jersey, New Mexico, New York, Oklahoma, Pennsylvania, Texas, Washington

Barcodes
04961333247974
Units affected
144 units
About this recall
Pentax Medical Video Processor; Model Number: EPK-i8020c;