Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Completed

Barco N.V. recalls MNA-420 ENC HDMI

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
Products
  • MNA-420 ENC HDMI
Company
Barco N.V.
Where it was sold

US: FL

States: Florida

Units affected
88 units
About this recall
MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display