Baxter Healthcare Corporation recalls Novum IQ LVP (Large Volume Pump)
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
- Products
-
- Novum IQ LVP (Large Volume Pump)
- Where it was sold
-
US distribution including Puerto Rico and OUS (International) Canada
States: Puerto Rico
- Barcodes
05413765851797- Units affected
- 43,922 units
- About this recall
- Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump