Hologic Inc recalls Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA System exceeded the limit for acceptability.
- Products
-
- Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A
- Company
- Hologic Inc
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of India, Australia, Spain, Estonia, Germany, Italy, Canada, Thailand, Dominican Republic, Portugal, Islas Canarias, Andorra, Switzerland, Taiwan, Austria, Korea, Republic of Philippines, Kuwait, Slovakia, China, Belgium, Ireland, United Kingdom, Malaysia, Palestinian Territory, Occupied, Peru, Singapore, Qatar, Algeria, Pakistan, New Zealand, Croatia, Brazil, Bangladesh, Saudi Arabia, Venezuela, Israel, Lebanon, Nepal, Romania, Greece, Hong Kong, Morocco, Indonesia, Cyprus, Jordan, France, Mexico, Colombia, Iran, Islamic, Republic of Chile, Turkey, Uruguay, Viet Nam, Serbia, Poland, Japan, Malta, Netherlands, South Africa.
Nationwide
- Barcodes
1542004550538415420045505698154200455058271542004550583415420045505865- Units affected
- 2,642 units (1037 units US; 1605 units OUS)
- About this recall
- Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A, Horizon-W, Horizon-WI, Horizon-C, Horizon-CI, Horizon-A-CN, Horizon-W-CN, Horizon-WI-CN, Horizon-CI-CN