Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Hologic Inc recalls Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA System exceeded the limit for acceptability.
Products
  • Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A
Where it was sold

Worldwide distribution - US Nationwide and the countries of India, Australia, Spain, Estonia, Germany, Italy, Canada, Thailand, Dominican Republic, Portugal, Islas Canarias, Andorra, Switzerland, Taiwan, Austria, Korea, Republic of Philippines, Kuwait, Slovakia, China, Belgium, Ireland, United Kingdom, Malaysia, Palestinian Territory, Occupied, Peru, Singapore, Qatar, Algeria, Pakistan, New Zealand, Croatia, Brazil, Bangladesh, Saudi Arabia, Venezuela, Israel, Lebanon, Nepal, Romania, Greece, Hong Kong, Morocco, Indonesia, Cyprus, Jordan, France, Mexico, Colombia, Iran, Islamic, Republic of Chile, Turkey, Uruguay, Viet Nam, Serbia, Poland, Japan, Malta, Netherlands, South Africa.

Nationwide

Barcodes
1542004550538415420045505698154200455058271542004550583415420045505865
Units affected
2,642 units (1037 units US; 1605 units OUS)
About this recall
Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A, Horizon-W, Horizon-WI, Horizon-C, Horizon-CI, Horizon-A-CN, Horizon-W-CN, Horizon-WI-CN, Horizon-CI-CN