Alcon Research LLC recalls Surgical Image Guidance Functionality
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.
- Products
-
- Surgical Image Guidance Functionality
- Company
- Alcon Research LLC
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI, PR and the countries of Canada, Japan, France.
Nationwide
- Barcodes
00380650003997- Units affected
- 230
- About this recall
- Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399