Beckman Coulter, Inc. recalls Dxl 9000 Access Immunoassay Analyzer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
Products
  • Dxl 9000 Access Immunoassay Analyzer
Where it was sold

Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.

States: Florida, Indiana, Kansas, New Hampshire, Pennsylvania, Texas

Barcodes
15099590732103
Units affected
143 devices
About this recall
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.