Baxter Healthcare Corporation recalls Spectrum IQ Infusion Pump
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
- Products
-
- Spectrum IQ Infusion Pump
- Where it was sold
-
US distribution to states of: AZ, FL
States: Arizona, Florida
- Barcodes
00085412610900- Units affected
- 6 units
- About this recall
- Spectrum IQ Infusion Pump, Product Code 3570009