ICU Medical Inc recalls Plum 360/A+ spare batteries. Item Number: SUB0000864.

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Products
  • Plum 360/A+ spare batteries. Item Number: SUB0000864.
Where it was sold

Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay.

Nationwide

Units affected
130,826 units