MEDLINE INDUSTRIES, LP - Northfield recalls The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professio

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.
Products
  • The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professio
Where it was sold

US Nationwide distribution tin the states of CA, FL, IA, and TX.

Nationwide

Barcodes
10195327584498101984592391681019845928165510198459282133101984590312121019845923362310195327257248
Units affected
1145
About this recall
The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit is provided sterile, single-use and includes a range of components dependent upon the specific procedure. These kits will be assembled using Medline and other manufacturers medical device components and drug components in some cases. Customers specify the kit contents, quantity, and placement of individual items in the kit.