mo-Vis BVBA recalls All-round Joystick R-net Light. Electrical wheelchair component.
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
- Products
-
- All-round Joystick R-net Light. Electrical wheelchair component.
- Company
- mo-Vis BVBA
- Where it was sold
-
US Nationwide distribution in the state of TX.
States: Texas
- Barcodes
05407008320096- Units affected
- 3 units
- About this recall
- All-round Joystick R-net Light. Electrical wheelchair component.