Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Products
  • AMERICAN CONTRACT SYSTEMS
Where it was sold

US distribution to states of: IL and LA

States: Illinois, Louisiana

Barcodes
001910722016210019107223535000191072235510001910722355270019107223337000191072238764
Units affected
1127 units
About this recall
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item Number LMCP22T; 4) TUR BASIN PACK, Model/Item Number LMTU31T; 5) CYSTO, Model/Item Number SACY80S; 6) UROLOGY PACK, Model/Item Number SPUR64F.