American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Products
-
- AMERICAN CONTRACT SYSTEMS
- Where it was sold
-
US distribution to states of: IL and LA
States: Illinois, Louisiana
- Barcodes
00191072203021001910722351760019107200370600191072005571001910720055880019107223726200191072137111001910722285120019107223527500191072103420001910720120980019107223554100191072205742001910722355890019107222259600191072233806- Units affected
- 6617 units
- About this recall
- AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item Number ANCV67X; 2) KIDNEY PERFUSION PACK - 208439, Model/Item Number ANKP15AD; 3) LAPCHOLAPPY PACK, Model/Item Number FYLA38G; 4) CYSTO PACK, Model/Item Number HSCY13M; 5) PICC DRESSING CHANGE KIT, Model/Item Number HSDR10K; 6) MIDLINE CHANGE KIT, Model/Item Number HSMD20H; 7) SHOULDER ARTHOSCOPY PACK, Model/Item Number PDSA12D; 8) PDA LF SJW, Model/Item Number SWPD12K; 9) AAA PACK - 206038, Model/Item Number UDAA63AH; 10) BASIC CAB PACK - 206043, Model/Item Number UDBC35AF; 11) BASIC HEART PACK - 206044, Model/Item Number UDBH37AX; 12) FREE FLAP DRAPE PACK (PS 907864), Model/Item Number UIFF72AD; 13) PROCUREMENT PACK, Model/Item Number UIPC89D; 14) THYROID PACK (PS 096552), Model/Item Number UITR45W; 15) TRAUMA PACK 175634, Model/Item Number UMTR17B; 16) TRAUMA PACK 175634, Model/Item Number UMTR17C.