American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Products
-
- AMERICAN CONTRACT SYSTEMS
- Where it was sold
-
US distribution to states of: IL and LA
States: Illinois, Louisiana
- Barcodes
0019107220724100191072235442- Units affected
- 204 units
- About this recall
- AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUMBER UISH55V.